Item | Specifications |
Detection Mode | Endpoint / Kinetics* / Optimum absorbance measurement*
Single / Dual wavelength measurement |
Photometry | Single light path, Top-irradiation/Bottom-reading (Posterior Dispersion) |
Measuring Range | -0.5 to 3.5 Abs |
Linearity | ±0.5%T or ±0.005Abs(0 to 2.5Abs at 492nm) |
Reproducibility | CV1.0% or less (Abs1.0, 492nm) |
Wavelength range | 340 to 750nm |
Measuring Direction | Horizontal or Vertical |
Plate Format | 96 well plate
100 microplate profiles are preset in data-processing PC application* |
Interference Filter | 450nm filter or a piece of filter among standard filters. |
Light Source | Halogen Lamp 6V-20W |
Mixing Function | Linear / 3-step intensity / time setting: 5 to 99.9 seconds |
Data Processing Software* | SF6 Software: Standard accessory (Windows XP, Mac-OSX(Power PC)
Regression formula 11 types / Cut off 2 types /
Data analysis formula 3 types / Kinetics 5 types |
Detector | Silicone photo diode |
External Port | Horizontal or Vertical |
Measuring Direction | USB for PC connection, D-SUB25 for optional external printer |
Outer Dimension | 30(W)×330(D)×165(H) mm |
Mass | Approx. 7.5kg |
Power Consumption | AC 100V to 240V, 50/60Hz, 55VA |
USB Port | Memory stick port (256MB to 2GB USB Memory are supported) |
Standard Filter | 405, 415, 450, 492, 530, 550, 570, 610, 630, 660nm |
Custom Filter | 340 to 750nm(in 1nm increments, except for standard filter) |
Optional Printer | DPU-414-30B-E, External thermal printer |