| Item | Specifications |
| Detection Mode | Endpoint / Kinetics* / Optimum absorbance measurement*
Single / Dual wavelength measurement |
| Photometry | Single light path, Top-irradiation/Bottom-reading (Posterior Dispersion) |
| Measuring Range | -0.5 to 3.5 Abs |
| Linearity | ±0.5%T or ±0.005Abs(0 to 2.5Abs at 492nm) |
| Reproducibility | CV1.0% or less (Abs1.0, 492nm) |
| Wavelength range | 340 to 750nm |
| Measuring Direction | Horizontal or Vertical |
| Plate Format | 96 well plate
100 microplate profiles are preset in data-processing PC application* |
| Interference Filter | 450nm filter or a piece of filter among standard filters. |
| Light Source | Halogen Lamp 6V-20W |
| Mixing Function | Linear / 3-step intensity / time setting: 5 to 99.9 seconds |
| Data Processing Software* | SF6 Software: Standard accessory (Windows XP, Mac-OSX(Power PC)
Regression formula 11 types / Cut off 2 types /
Data analysis formula 3 types / Kinetics 5 types |
| Detector | Silicone photo diode |
| External Port | Horizontal or Vertical |
| Measuring Direction | USB for PC connection, D-SUB25 for optional external printer |
| Outer Dimension | 30(W)×330(D)×165(H) mm |
| Mass | Approx. 7.5kg |
| Power Consumption | AC 100V to 240V, 50/60Hz, 55VA |
| USB Port | Memory stick port (256MB to 2GB USB Memory are supported) |
| Standard Filter | 405, 415, 450, 492, 530, 550, 570, 610, 630, 660nm |
| Custom Filter | 340 to 750nm(in 1nm increments, except for standard filter) |
| Optional Printer | DPU-414-30B-E, External thermal printer |