Item Model | Specifications | Others |
MTP-810Lab | MTP-880Lab |
Detecting method |
| Measuring mode | End point/Kinetic/Optimal value measurement(Abs only)/Multi-point | |
FL measurement | 2Ex-2Em / 1Ex-1Em / 2Ex-1Em / 1Ex-2Em Time Resolved Fluorescence Measurement |
CL measurement | All-wavelength | All-wavelength/1-wavelength /2-wavelength |
Abs measurement | Single wavelength / Dual wavelength |
Wavelength range |
| Normal FL | From 300 to 750nm (Ex: from 260 to 710nm) | |
TRF | 300 to 750nm (Ex:260 to 710nm) | 400 to 850nm (Ex: 260 to 710nm) |
CL | 300 to 750nm | 400 to 850nm |
Abs | 260 to 999nm |
Abs measurement |
| Range | -0.5 to 3.5Abs | |
Linearity | 0 to 2.5Abs (±0.5%T or ±0.005Abs within 0 to 2.5Abs/492nm/96-well Plate) |
Filter |
| Disk 1 (EX/λ) | 450, 492(Abs) 365, 490nm(Ex) (Max. 3 added filters) | |
Disk 2 (EM) | 450, 530, (615 for TRF)nm (Max. 3 added filters) | 450, 530 nm (Max. 5 added filters) |
Disk 3 (EM/CL) | N/A | 615nm for TRF(Max. 2 filters) |
Plate Format |
FL/TRF/CL | : | 6, 12, 24, 48, 96, 384 well (flat-black) |
Abs | : | 6, 12, 24, 48, 96, 384well (flat-clear)
| *Custom plate operational (Height of plate should be 21mm or less.) |
| 258 kinds of Plate profile are preset |
Measurement time | 50 sec. / 96 well plate (at minimum) | |
Times & interval of measurement | 999 Sec. 999 times (Kinetic Measurement) | |
Stirrer Function |
| Mode | Rotation/Linear, (3-step Level) | |
Time | From 5 to 99.9 sec. |
Thermal Control Function | Thermal range from 5 to 45ºC (1ºC step adjustable) | *Optional |
Multi point reading | 9 to 81 points can be read at a measurement. | PC Required |
Dispensing function | Independent 2 reagents dispense, Dispense volume : 20 to 200µL Dispense precision: CV 1% or less (6 to 96 well-plate are available) *Variable dispense volume function for each well, Full-time reading function | *Optional |
Light Source | Xenon flash lamp | |
Detector |
| FL | Photomultiplier | |
TRF/CL | Photomultiplier | Photon Counting Photomultiplier |
Abs | Silicon photo-diode |
Power | AC100V(50/60Hz) 200VA | |
Dimension and weight | 420(W)×450(D)×315(H)mm 32kg/33.5kg (MTP-810Lab/880Lab) | |
Software |
| Software SF6 is bundled as standard accessory.
Windows 2000/XP/Vista and Mac OSX (PowerPC) are supported. |
| Main function
1) | PC Control function |
2) | Data Processing function: 11 Regression models, Two Cut-off models /Three Data analysis models / Five Kinetic models |
3) | Controlling function F Instrumental operation logging, Function of Recording communication log between Instrument and PC |
|
| PC Required |